FDA Adverse Event Malfunction Summary report: N

STANDOP VOLISTA®

MDR report key: 21467580 · Received February 26, 2025

Report

Report Number
9710055-2025-0000082
Event Type
Malfunction
Date Received
February 26, 2025
Date of Event
February 7, 2025
Report Date
February 26, 2025
Manufacturer
MAQUET SAS
Product Code
FSY
UDI-DI
03700712401429
PMA / PMN Number
K130513
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. IT WAS STATED AND CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SPRING ARM DUST COVER (UPPER SIDE OF HORIZONTAL ARM) FELL OFF. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION. BASED ON THE INFORMATION COLLECTED, IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE SURGICAL LIGHTS DID NOT MEET ITS SPECIFICATION, SINCE SPRING ARM METAL COVER DETACHMENT COULD BE CONSIDERED AS TECHNICAL DEFICIENCIES, AND IN THIS WAY THE DEVICE CONTRIBUTED TO EVENT. IT IS UNKNOWN IF CLAIMED DEVICES WERE OR WERE NOT BEING USED FOR PATIENT TREATMENT OR DIAGNOSIS WHEN THE EVENT TOOK PLACE. ACCORDING TO SUBJECT MATTER EXPERT AT MANUFACTURER¿S, THE MOST PROBABLE ROOT CAUSES ARE : NON-CONFORMITY OF THE METAL COVERS ASSEMBLY. DEGRADATION OF THE METAL COVERS. IMPROPER USE (COLLISION WITH ANOTHER DEVICE). MAQUET SAS ANALYSIS SHOWS THAT THE METAL STRIP COMES OUT OF THE COVERS WHEN IT IS NOT CLIPPED PROPERLY. IN THE SCOPE OF OUR CONTINUOUS IMPROVEMENT POLICY, MAQUET SAS INITIATED A MODIFICATION FILE (B)(4) TO INCLUDE THIS DUST COVER FITTING PROCEDURE IN THE TECHNICAL DOCUMENTATIONS WITH ALL SPRING ARMS. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DOES NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.

Description of Event or Problem · 0

GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. IT WAS STATED AND CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SPRING ARM DUST COVER (UPPER SIDE OF HORIZONTAL ARM) FELL OFF. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1248791 STANDOP VOLISTA® LIGHT, SURGICAL, CEILING MOUNTED FSY MAQUET SAS ARDVST229011A 03700712401429

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown