FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21458880 · Received February 25, 2025

Report

Report Number
2955842-2025-01994
Event Type
Malfunction
Date Received
February 25, 2025
Date of Event
January 29, 2025
Report Date
January 29, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110720
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE CONDUCTED A TEST DRIVE OF THE SYSTEM AND COULD NOT REPRODUCE THE ISSUE. THE SYSTEM WAS VERIFIED AND READY TO USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ENDOMETRIOSIS RESECTION SURGICAL PROCEDURE, CUSTOMER CONTACTED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) AND REPORTED UNIVERSAL SURGICAL MANIPULATOR (USM) #2, LED WAS YELLOW AND COULD NOT ADVANCE WELL. IT WAS STATED THAT THE CASE WAS COMPLETED BY SWITCHING TO USM #4 INSTEAD. TSE VIEWED LOGS BUT DID NOT FIND ANY RELATED ERRORS. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PORTS WERE PLACED PRIOR TO ISSUE BEING IDENTIFIED. SURGEON ELECTED NOT TO USE ARM #2 (NON-WORKING ARM) IN ORDER TO PROCEED WITH THE CASE. THE PROCEDURE WAS COMPLETED WITH 3-ARMS USED DUE TO ARM #2 NOT COMMUNICATING WELL WITH SURGEON'S CONSOLE AND THE LIGHT ON ARM #2 WAS ORANGE. ALSO, THE PHYSICIAN ASSISTANT COULD NOT ADVANCE ROBOTIC ARM INTO THE TROCAR. PROCEDURE WAS COMPLETED ROBOTICALLY WITH ARMS #1, 3, AND 4. THERE WAS NO INJURY TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411538 DAVINCI XI PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380652-55 N/A 00886874110720

Patients

Seq Age Sex Outcome Treatment
1 NA Female DA VINCI INSTRUMENTS AND ACCESSORIES