ACCESS 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-01461
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- July 23, 2010
- Report Date
- July 24, 2007
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- MMI
- PMA / PMN Number
- K021814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
HETEROPHILE INTERFERENCE IS THE ROOT CAUSE OF THE EVENT. THE FIELD SERVICE ENGINEER (FSE) REPLACED THE ASPIRATE PROBES, PERI-PUMP TUBING, WASH ARM LEAD SCREW, MIXER AND MIXER ROLLER ASSEMBLIES, MIXER BELT, AND THE SUBSTRATE PROBE. THE FSE SUCCESSFULLY COMPLETED SYSTEM CHECK AND PERFORMED HS LUMWASH SON-INC TEST WHICH MET SPECIFICATIONS. THE RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS, AND THE INSTRUMENT WAS VERIFIED PER ESTABLISHED PROCEDURES. PRODUCT LABELING: FOR ASSAYS EMPLOYING ANTIBODIES, THE POSSIBILITY EXISTS FOR INTERFERENCE BY HETEROPHILE ANTIBODIES IN THE PT SAMPLE. PTS WHO HAVE BEEN REGULARLY EXPOSED TO ANIMALS OR HAVE RECEIVED IMMUNOTHERAPY OR DIAGNOSTIC PROCEDURES UTILIZING IMMUNOGLOBULINS OR IMMUNOGLOBULIN FRAGMENTS MAY PRODUCE ANTIBODIES, E.G. HUMAN ANTI-MOUSE ANTIBODIES (HAMA), THAT INTERFERE WITH IMMUNOASSAYS. ADDITIONALLY, OTHER HETEROPHILE ANTIBODIES SUCH AS HUMAN ANTI-GOAT ANTIBODIES MAY BE PRESENT IN PT SAMPLES. SUCH INTERFERING ANTIBODIES MAY CAUSE ERRONEOUS RESULTS. CAREFULLY EVALUATE THE RESULTS OF PTS SUSPECTED OF HAVING THESE ANTIBODIES. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS CONDUCTED BETWEEN (B)(4) 2008 TO (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS. RELATED MDRS: 2122870-2011-01459 AND 2122870-2011-01460. THE CUSTOMER COLLECTS SAMPLES IN 12X75 PLASTIC PLASMA SEPARATOR TUBE (PST) SODIUM HEPARIN TUBES WITH GEL. THE CUSTOMER CENTRIFUGES SAMPLES AT 8500 RPM FOR 3 MINUTES IN A STATSPIN CENTRIFUGE. THE SPECIMENS ARE SAMPLED FROM THE PRIMARY TUBES. THE CUSTOMER SENT ONE SERUM AND ONE PLASMA SAMPLE FROM THE PT TO BECKMAN COULTER, INC. FOR FURTHER ANALYSIS. QUALITY CONTROL (QC) WAS WITHIN SPECIFICATIONS DURING THE TIME FRAME OF THIS EVENT. THE CUSTOMER PERFORMED A SYSTEM CHECK WHICH PRODUCED FAILING WASHED AND UNWASHED PERCENT VALUES. THE CUSTOMER DID NOT SUPPLY THE ACTUAL SYSTEM CHECK DATA PRINT-OUTS. BECKMAN COULTER, INC. CONFIRMED THE CUSTOMER'S RESULTS ON THE NEAT SAMPLE WITH A TROPONIN (ATNI) VALUE OF 7.615 NG/ML. THE ATNI CONCENTRATION SIGNIFICANTLY DECREASED (BY 99%) AFTER THE ADDITION OF A HETEROPHILE BLOCKER. IT IS CONCLUSIVE THAT THE CUSTOMER'S RESULTS ARE FALSELY ELEVATED DUE TO INTERFERING SUBSTANCES IN THE PT BLOOD. THE OCCURRENCE OF SUCH INTERFERENCE IS EXTREMELY LOW. AS ALL MANUFACTURERS, BECKMAN COULTER INC. ADDS ADDITIVES TO PRODUCTS TO LIMIT SUCH AN EFFECT. BUT IN SOME RARE CASES, INTERFERENCE IS SUCH THAT ADDITIVE PRODUCTS ARE NOT EFFICIENT ENOUGH.
THE CUSTOMER ALLEGED HIGH ERRONEOUS TROPONIN ACCUTNI RESULTS ABOVE THE ACUTE MYOCARDIAL INFARCTION (AMI) CUTOFF FOR ONE PT'S SAMPLES INVOLVING ACCESS 2 IMMUNOASSAY SYSTEM. THIS REPORT REFERS TO SAMPLE NUMBER THREE. THE CUSTOMER SENT SPECIMENS FROM THE PT TO BECKMAN COULTER, INC. FOR FURTHER ANALYSIS AND HETEROPHILE INTERFERENCE WAS CONFIRMED. IT IS UNK IF THE ERRONEOUS RESULTS WERE RELEASED OUT OF THE LABORATORY. THERE WAS NO REPORT OF PT INJURY OR CHANGE IN PT TREATMENT ASSOCIATED WITH THIS EVENT. THE FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY AND ASSESSED THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS 2 IMMUNOASSAY SYSTEM | MMI | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | ACCUTNI ON THE ACCESS IMMUNOASSAY SYSTEMS |