FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 21451651 · Received February 24, 2025

Report

Report Number
9617229-2025-03170
Event Type
Injury
Date Received
February 24, 2025
Date of Event
January 31, 2025
Report Date
May 19, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9 H3 H6. LABORATORY ANALYSIS SUMMARY THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON APRIL 03, 2025 WITH LOT NUMBER 3230563. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: DEFLATION: OBSERVED AN OPENING ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE, CREASES AND WEAR ABRASION WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION: DEFLATION.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: A4, B5, D6B, H6.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. HOLE IN RADIUS (EDGE) OBSERVED DURING SURGERY. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAIN IMPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL LATER REPORTED "HOLE IN RADIUS (EDGE)." DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
521304 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 3230563

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female Required Intervention