FDA Adverse Event Injury Summary report: N

BECKMAN COULTER LAB AUTOMATION SYSTEM

MDR report key: 2144909 · Received June 30, 2011

Report

Report Number
2050012-2011-02574
Event Type
Injury
Date Received
June 30, 2011
Date of Event
June 2, 2011
Report Date
June 2, 2011
Manufacturer
BECKMAN COULTER BIOMEDICAL GMBH
Product Code
JQP
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED FOR THIS EVENT. THE REQUIRED SECOND PAIR OF GLOVES WAS NOT WORN DURING MAINTENANCE. THE OMITTED COTTON GLOVES ARE NOT FOR PROTECTION PURPOSE BUT FOR AVOIDING TRACES OF GREASE OR FAT ON THE MIRROR. AS THE FIELD SERVICE TECHNICIAN WAS NOT WEARING THE REQUIRED PPE, USE ERROR IS A CONTRIBUTING CAUSE TO THIS EVENT.

Description of Event or Problem · 1

A BECKMAN COULTER FIELD SERVICE ENGINEER REPORTED THAT DURING ROUTINE SERVICE AND CLEANING OF THE BECKMAN COULTER LAB AUTOMATION SYSTEM, HE GRAZED HIS HAND AGAINST THE EDGE OF ONE OF THE MIRRORS USED IN THE TUBE INSPECTION UNIT (TIU). THIS AREA IS ONLY ACCESSIBLE TO MAINTENANCE PERSONNEL AND THE MAINTENANCE INVOLVED IN THIS EVENT IS ONLY PERFORMED BY BECKMAN COULTER FIELD SERVICE ENGINEERS. THE MIRROR THAT CAUSED THE GRAZE IS CHAMFERED BY SPECIFICATIONS, BUT IT IS NOT FRAMED OR ROUNDED. CONTACT WITH THE MIRROR RESULTED IN A CUT TO HIS FINGER WHICH BLED SLIGHTLY. THE MIRROR INVOLVED HAD BEEN DISINFECTED WITH ETHANOL JUST PRIOR TO THE INJURY ASSOCIATED WITH THIS REPORT. THE SERVICE ENGINEER WAS WEARING A LAB COAT AND RUBBER GLOVES AT THE TIME OF THE INCIDENT, HOWEVER, WAS NOT WEARING THE LINT-FREE COTTON GLOVES REQUIRED IN THE USER'S GUIDE. THE FSE WENT TO THE MEDICAL DEPARTMENT AND A BLOOD SAMPLE WAS TAKEN AS A BASELINE FOR IN-VITRO DIAGNOSTIC TESTING FOR BLOOD-BORNE DISEASES. BASELINE TESTING SHOWED NO (B)(4), (B)(4) OR (B)(4). THE FSE RECEIVED A PROACTIVE TETANUS INOCULATION, BUT NO OTHER PROPHYLAXIS WAS ADMINISTERED. THIS EVENT OCCURRED AT: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN COULTER LAB AUTOMATION SYSTEM LAB AUTOMATION JQP BECKMAN COULTER BIOMEDICAL GMBH NA NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention