FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 21446037 · Received February 24, 2025

Report

Report Number
3008642652-2025-01962
Event Type
Malfunction
Date Received
February 24, 2025
Date of Event
February 18, 2025
Report Date
February 24, 2025
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER ON) WAS CONFIRMED DUE TO A DEFECTIVE U104 PXA PROCESSOR CHIP ON THE COMPUTER/ANALOG BOARD, CAUSING THE COMMUNICATION ERROR. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. THERE WAS NO ADVERSE EVENT THAT RESULTED FROM THE DAMAGED MONITOR.

Description of Event or Problem · 0

A US DISTRIBUTOR REPORTED THAT A MONITOR WILL NOT POWER ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
557966 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM 00855778005005

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown