FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 21446037
·
Received February 24, 2025
Report
- Report Number
- 3008642652-2025-01962
- Event Type
- Malfunction
- Date Received
- February 24, 2025
- Date of Event
- February 18, 2025
- Report Date
- February 24, 2025
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005005
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER ON) WAS CONFIRMED DUE TO A DEFECTIVE U104 PXA PROCESSOR CHIP ON THE COMPUTER/ANALOG BOARD, CAUSING THE COMMUNICATION ERROR. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. THERE WAS NO ADVERSE EVENT THAT RESULTED FROM THE DAMAGED MONITOR.
Description of Event or Problem · 0
A US DISTRIBUTOR REPORTED THAT A MONITOR WILL NOT POWER ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557966 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | LIFEVEST WCD 4000 SYSTEM | 00855778005005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |