FDA Adverse Event Malfunction Summary report: N

NEEDLE 18X1-1/2 BLUNT FILL

MDR report key: 21445515 · Received February 24, 2025

Report

Report Number
1911916-2025-00139
Event Type
Malfunction
Date Received
February 24, 2025
Date of Event
February 15, 2025
Report Date
March 10, 2025
Manufacturer
BECTON DICKINSON
Product Code
GAA
UDI-DI
30382903051800
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP. IT WAS REPORTED THE NEEDLE CORED THE RUBBER STOPPER. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305180 AND LOT NUMBER 4212676. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Description of Event or Problem · 0

MATERIAL #305180. LOT #4212676. VERBATIM: ON 3 SEPARATE OCCASIONS, I DREW UP SOME MEDS USING A RED, BLUNT FILL NON-CORING NEEDLE AND DREW UP WHAT LOOKED LIKE RUBBER STOPPER INTO MY MED. I HAVE NEVER HAD THIS HAPPEN TO ME PRIOR TO THIS EVENT. NO PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
557935 NEEDLE 18X1-1/2 BLUNT FILL NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE GAA BECTON DICKINSON 4212676 30382903051800

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown