FDA Adverse Event Malfunction Summary report: N

ZOOM CRITICAL CARE BED

MDR report key: 2143721 · Received June 3, 2011

Report

Report Number
1831750-2011-05454
Event Type
Malfunction
Date Received
June 3, 2011
Date of Event
May 12, 2011
Report Date
May 12, 2011
Manufacturer
STRYKER MEDICAL
Product Code
FNL
Removal / Correction Number
Z-2587-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULTS: ACTUATOR BOX.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE FOWLER WOULD NOT LOWER NEITHER ELECTRONICALLY NOR VIA THE CPR RELEASE FUNCTION. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES ARE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZOOM CRITICAL CARE BED A/C POWERED HOSPITAL BED FNL STRYKER MEDICAL 2040 NA

Patients

Seq Age Sex Outcome Treatment
1