FDA Adverse Event Summary report: N

GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN

MDR report key: 2143474 · Received March 16, 2011

Report

Report Number
3003900188-2011-00011
Date Received
March 16, 2011
Date of Event
January 29, 2011
Manufacturer
LINDE GAS NORTH AMERICA LLC
Product Code
ECX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION IS ONGOING. EVAL: ACCIDENTAL IGNITION AND FIRE IN OXYGEN VALVE.

Description of Event or Problem · 1

THIS IS A SPONTANEOUS MEDICAL INCIDENT REPORT FROM A MAINTENANCE ENGINEER AT A HOSP IN BRAZIL CONCERNING A (B)(6) MALE PORTER AT THE HOSP. ON (B)(6) 2011 WHEN THE PORTER TRIED TO OPEN THE COMBILITE TYPE (B)(4) CYLINDER VALVE (GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN) OF AN OXYGEN CYLINDER USED FOR TRANSPORTATION A FIRE STARTED AND BURNT HIS HAND. THE PORTER HAD 1ST AND 2ND DEGREE BURNS AND WAS ASSISTED IN THE HOSP WHERE HE WORKS AND WAS SENT HOME FOR 5 WORKING DAYS. TREATMENT GIVEN FOR THE BURNS WERE ASEPTIC AND DRESSING WITH SILVER SULFADIAZINE. THEREAFTER HE WAS RECOVERED AND IS BACK TO WORK IN THE SAME ACTIVITIES AS BEFORE THE INCIDENT. ADD'L INFO IS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN NONE ECX LINDE GAS NORTH AMERICA LLC 0710160 4195896

Patients

Seq Age Sex Outcome Treatment
1 27 YR