FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2143437 · Received June 29, 2011

Report

Report Number
2531779-2011-04555
Event Type
Injury
Date Received
June 29, 2011
Date of Event
May 31, 2011
Report Date
May 31, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER, THE PATIENT'S DAUGHTER, REPORTED THAT ON (B)(6) 2011 THE PATIENT NOTED A LOSS OF PRIME ON THE INSULIN PUMP, AND AFTERWARDS EXPERIENCED "AN INSULIN REACTION". THE PATIENT'S SPEECH WAS SLURRED AND SHE EXPERIENCED CONFUSION. THE PATIENT DID NOT TEST HER BLOOD GLUCOSE LEVEL AT THAT TIME. THE PATIENT ADMINISTERED SELF-TREATMENT BY CONSUMING A CANDY BAR; HER SUBSEQUENT BLOOD GLUCOSE READING WAS 67 MG/DL AT 6:57 PM. SHE DID NOT SEEK ANY MEDICAL ATTENTION. TROUBLESHOOTING REVEALED THAT THE PATIENT HAD REMAINED CONNECTED TO THE PUMP WHEN SHE ADDRESSED THE SECOND LOSS OF PRIME ISSUE, WHICH CAN CONTRIBUTE TO OVER-INFUSION OF INSULIN. THE TECHNICAL SUPPORT REPRESENTATIVE ALSO DETERMINED THE BATTERY CAP WAS WORN AND WOULD NOT REMAIN SECURE, AND THAT THE PATIENT HAD NOT REPLACED IT. THE PATIENT'S TECHNIQUE WAS INCORRECT IN THAT SHE PRIMED THE PUMP WHILE IT WAS ATTACHED TO HER BODY AND SHE DID NOT REPLACE THE WORN BATTERY CAP. HOWEVER, AS THE PATIENT ALLEGEDLY SUFFERED SYMPTOMS SUGGESTING SEVERE HYPOGLYCEMIA AFTERWARDS, WHILE USING THE PUMP, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 56 YR Life Threatening