FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2143415 · Received June 29, 2011

Report

Report Number
2050012-2011-02517
Event Type
Malfunction
Date Received
June 29, 2011
Date of Event
June 3, 2011
Report Date
June 3, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011. THE FSE PERFORMED TROUBLESHOOTING, SERVICED THE INSTRUMENT AND NO DRIPPING SEEN BY FSE. ON (B)(6) 2011, CUSTOMER REPORTED TO FSE THAT LEAKING CONTINUED. THE FSE REPLACED THE A/B ASSEMBLY; HOWEVER, LEAKING WAS STILL OBSERVED BY FSE. THE FSE EXAMINED INSTRUMENT FURTHER AND FOUND THE VACUUM VALVE WAS NOT OPENING FOR THE REAGENT PROBE. THE FSE TOOK VALVE APART AND CLEANED. NO FURTHER LEAKING OBSERVED BY FSE. REPAIR VERIFIED PER ESTABLISHED PROCEDURES. AN MDR NUMBER: 2050012-2011-02516 WAS SUBMITTED IN ASSOCIATION WITH A LEAK ON THIS INSTRUMENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) AND REPORTED THAT THEY CONTINUED TO GET CUVETTES FAILING WATER BLANK TEST AND OBSERVED LEAKING FROM REAGENT PROBES. THE ISSUES WERE ASSOCIATED WITH THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM. A CUSTOMER TECHNICAL SUPPORT (CTS) GENERATED A SERVICE REQUEST. NO INJURY WAS REPORTED ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1