FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2143394 · Received June 29, 2011

Report

Report Number
2050012-2011-02516
Event Type
Malfunction
Date Received
June 29, 2011
Date of Event
May 29, 2011
Report Date
May 29, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011. THE FSE REPLACED REAGENT SYRINGE AND T-VALVE; PERFORMED ALIGNMENTS AND RAN QC. THE FSE ORDERED A BUBBLE GENERATOR IN CASE DRIPPING RETURNED. ON (B)(4) 2011, THE LAB REPORTED TO FSE THAT THE INSTRUMENT CONTINUED TO LEAK. THE FSE RETURNED TO THE CUSTOMER'S LAB ON (B)(4) 2011 AND REPLACED THE BUBBLE GENERATOR, PRIMED THE ANALYZER AND RAN QC. CUSTOMER WAS TO CONTINUE TO MONITOR.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) AND REPORTED THAT THEY WERE EXPERIENCING ONGOING CUVETTE WATER BLANK ERRORS ON UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM. A CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER IN TROUBLESHOOTING. CUSTOMER CHECKED CARTRIDGE CHEMISTRIES (CC) TOP END AND OBSERVED EVIDENCE OF LEAKING FROM THE REAGENT PROBES. CUSTOMER CHECKED PROBES/FITTINGS AND REAGENT SYRINGE; ALL WERE TIGHT. CUSTOMER PERFORMED QC ON THE CC SIDE WITH GOOD RESULTS. CTS GENERATED A SERVICE REQUEST TO CHECK THE INSTRUMENT. NO INJURY WAS REPORTED ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1