FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2143393 · Received June 29, 2011

Report

Report Number
2531779-2011-04551
Event Type
Malfunction
Date Received
June 29, 2011
Report Date
May 30, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP DATE AND TIME REVERTED TO DEFAULT SETTINGS AFTER THE BATTERY WAS REPLACED. THE PATIENT RESET THE CORRECT DATE/TIME AFTER REPLACING THE BATTERY. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 36 YR