FDA Adverse Event
Malfunction
Summary report: N
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
MDR report key: 2143392
·
Received June 29, 2011
Report
- Report Number
- 1030489-2011-00826
- Event Type
- Malfunction
- Date Received
- June 29, 2011
- Date of Event
- May 31, 2011
- Report Date
- May 31, 2011
- Manufacturer
- MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE INFERIOR SHAFT IS BROKEN OFF AT THE CENTER OF THE JAW PIVOT PIN FORWARD. THE BROKEN OFF PORTION OF THE SHAFT IS MISSING AND WAS NOT RETURNED FOR ANALYSIS. MICROSCOPIC EXAMINATION DISCOVERED A FAIRLY BRITTLE FRACTURE. THE LOCATION, DIRECTION, AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFT IS CONSISTENT WITH APPLICATION OF EXCESSIVE FORCE, RESULTING IN THE FOREGOING EVENT. DEVICE HISTORY RECORDS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT BROKE DURING SURGERY, NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING | NA | GZ05M041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |