RADIAL JAW 4 BIOPSY FORCEPS
Report
- Report Number
- 3005099803-2011-02255
- Event Type
- Malfunction
- Date Received
- June 29, 2011
- Date of Event
- June 9, 2011
- Report Date
- June 9, 2011
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA
- Product Code
- FCL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ZA
- Reporter Occupation
- PHYSICIAN
Narratives
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE NEEDLE TAIL WAS BENT. ADDITIONALLY, NO ABNORMALITIES WERE NOTED WITH THE DEVICE RIVETING AND WELDING WHICH WERE WITHIN DESIGN SPECIFICATIONS. FUNCTIONALLY THE DEVICE JAWS WOULD OPEN AND CLOSE NORMALLY CONSIDERING THE BENT NEEDLE TAIL. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT SINCE THE RETURNED DEVICE VISUALLY REFLECTS THE REPORTED ISSUE. THE PULL WIRES OF THE DEVICE WERE INTACT AND WERE NOT BROKEN. SINCE THERE ARE CONTROLS IN THE MANUFACTURING PROCESS THAT VERIFY PRODUCT INTEGRITY, THE SPECIFIC CAUSE OF THE FAILURE CANNOT BE IDENTIFIED. THEREFORE THE MOST PROBABLE ROOT CAUSE IS UNDETERMINABLE, HOWEVER, AN INVESTIGATION IS UNDERWAY TO ADDRESS THIS ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. (B)(4).
THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 4 SINGLE USE BIOPSY FORCEPS LARGE CAPACITY DEVICE WAS TO BE USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, AFTER REMOVAL FROM ITS PACKAGING, THE DEVICE WAS INSPECTED AND THE PULL WIRES WERE FOUND TO BE DAMAGED AND LOOSE. THE PROCEDURE WAS COMPLETED WITH ANOTHER RADIAL JAW 4 SINGLE USE BIOPSY FORCEPS LARGE CAPACITY DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 4 SINGLE USE BIOPSY FORCEPS LARGE CAPACITY DEVICE WAS TO BE USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, AFTER REMOVAL FROM ITS PACKAGING, THE DEVICE WAS INSPECTED AND THE PULL WIRES WERE FOUND TO BE DAMAGED AND LOOSE. THE PROCEDURE WAS COMPLETED WITH ANOTHER RADIAL JAW 4 SINGLE USE BIOPSY FORCEPS LARGE CAPACITY DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIAL JAW 4 BIOPSY FORCEPS | FORCEPS, BIOPSY, NON-ELECTRIC | FCL | BOSTON SCIENTIFIC - COSTA RICA | M00513331 | 0013937795 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR |