FDA Adverse Event Malfunction Summary report: N

THORATEC® HEARTMATE 3® LVAS IMPLANT KIT

MDR report key: 21433501 · Received February 21, 2025

Report

Report Number
2916596-2025-01273
Event Type
Malfunction
Date Received
February 21, 2025
Date of Event
August 22, 2023
Report Date
April 29, 2025
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
UDI-DI
00813024013297
PMA / PMN Number
P160054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF OUTFLOW GRAFT OBSTRUCTION (OGO) WAS UNABLE TO BE CONFIRMED AS NO IMAGES WERE SUBMITTED FOR REVIEW AND THE DEVICE REMAINS IN USE. A SPECIFIC CAUSE FOR THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED THROUGH THIS EVALUATION. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE CUSTOMER REGARDING THE EVENT; HOWEVER, NO ADDITIONAL INFORMATION WAS PROVIDED. THE PATIENT REMAINS ONGOING ON HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM, SERIAL NUMBER (B)(6). THE RELEVANT SECTIONS OF THE DEVICE HISTORY RECORDS FOR (B)(6) WERE REVIEWED AND SHOWED NO DEVIATIONS FROM MANUFACTURING OR QUALITY ASSURANCE SPECIFICATIONS. THE CURRENT REVISIONS OF THE INSTRUCTIONS FOR USE (IFU) AND PATIENT HANDBOOK CAN BE FOUND ON THE EIFU PAGE OF THE ABBOTT WEBSITE. THE HEARTMATE 3 LVAS IFU IS CURRENTLY AVAILABLE. SECTION 5 OF THE IFU, ¿SURGICAL PROCEDURES¿, OUTLINES CONSIDERATIONS FOR PUMP PLACEMENT AND PROVIDES INSTRUCTIONS REGARDING THE PREPARATION, INSTALLATION, AND ORIENTATION OF THE SEALED OUTFLOW GRAFT. SECTION 5, UNDER "ATTACHING THE SEALED OUTFLOW GRAFT TO THE PUMP", INSTRUCTS THE USER TO VERIFY THAT THE OUTFLOW GRAFT IS NOT TWISTED OR KINKED BY CHECKING THE POSITION OF THE BLACK LINE ON THE GRAFT ABOVE AND BELOW THE BEND RELIEF AND ENSURING THAT THE LINE IS STRAIGHT. SECTION 5, UNDER "DE-AIRING THE PUMP", CAUTIONS THE USER: "DO NOT ROTATE/TWIST THE SEALED GRAFT. CHECK THE ALIGNMENT OF THE BLACK LINE ON THE GRAFT TO VERIFY THAT THE SEALED GRAFT IS NOT TWISTED OR KINKED." SECTION 5, UNDER "PREIMPLANT PROCEDURES" AND "IMPLANT PROCEDURES" CAUTIONS THE USER: "THE SEALED OUTFLOW GRAFT MUST NOT BE KINKED OR POSITIONED WHERE IT COULD ABRADE AGAINST A PUMP COMPONENT OR BODY STRUCTURE". NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.

Additional Manufacturer Narrative · 0

NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ONCE THE MANUFACTURER¿S INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT HAD AN OUTFLOW GRAFT OBSTRUCTION (OGO) REPAIR ON (B)(6) 2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1666744 THORATEC® HEARTMATE 3® LVAS IMPLANT KIT VENTRICULAR (ASSIST) BYPASS DSQ THORATEC CORPORATION 106524US 6957202 00813024013297

Patients

Seq Age Sex Outcome Treatment
1 77 YR Male