FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 ARTICULATING

MDR report key: 2143275 · Received June 28, 2011

Report

Report Number
3005075853-2011-02643
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
May 27, 2011
Report Date
May 31, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. HOWEVER, AFTER FURTHER ANALYSIS THE KNIFE WAS NOTED TO BE NICKED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE TO THE KNIFE IS WHEN THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE AND INTERFERING WITH THE KNIFE PATH DURING DEVICE FIRING, PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. ADDITIONALLY, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CLOSING COLOSTOMY PROCEDURE, THE LINE OF STAPLES DIDN'T FORM COMPLETELY. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL FOLLOWUP: ON WHAT TISSUE TYPE WAS THE DEVICE USED? AT WHAT LOCATION ON THE TISSUE? DESCENDING COLON, THICK TISSUE. WAS THE DEVICE FIRED ON TISSUE THAT HAD BEEN RADIATED OR HAS THE PATIENT BEEN TAKING SYSTEMIC STEROIDS? NO. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC)? 2ST. IF ECHELON, DURING WHICH STROKE DID THE EVENT OCCUR? 1ST. WHAT COLOR CARTRIDGE WAS BEING USED? GOLD. WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? BEFORE GREEN AND AFTER GOLD. WAS BUTTRESSING MATERIAL UTILIZED? IF SO, WHICH PRODUCT? NO. WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? NO. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? NO. WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? FIRING. AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? ALL RETURNED AUTOMATICALLY TO THE ORIGINAL POSITION. WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? YES, THE TISSUE WAS INTO THE JAWS AND THEY HAD GREAT DIFFICULTY IN REMOVING THE TISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 ARTICULATING STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK H43J0G

Patients

Seq Age Sex Outcome Treatment
1