FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2143249 · Received June 28, 2011

Report

Report Number
2531779-2011-04538
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE INVESTIGATION RESULTS COMPLETED ON 05/20/2011. CORRECTION #: 2531779-03/24/2010-003-R. DURING EVALUATION, THE FORCE SENSOR FLEX CONNECTOR WAS FOUND TO BE TORN AT THE FORCE SENSOR PINS AND THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED. THE FORCE SENSOR WAS FOUND TO HAVE A BAD CALIBRATION.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED AN INTERMITTENT FLEX CONNECTION AT THE FORCE SENSOR PINS AND A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 46 YR