FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM

MDR report key: 2143217 · Received June 28, 2011

Report

Report Number
2134265-2011-02374
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
March 24, 2011
Report Date
March 29, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SG
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION FOUND THE DEVICE WAS RETURNED TO THE WITH STENT DAMAGE. STRUT ROWS IN THE MIDDLE SECTION OF THE STENT WERE RAISED AND MISALIGNED. THIS TYPE OF DAMAGE IS CONSISTENT WITH THE STENT MEETING RESISTANCE UPON ADVANCEMENT AND/OR WITHDRAWAL. THERE WAS A KINK IN THE LUMEN 170MM DISTAL TO THE TIP. THERE WERE ALSO KINKS IDENTIFIED ALONG THE ENTIRE LENGTH OF THE HYPOTUBE SHAFT. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE BEING APPLIED TO THE DELIVERY SYSTEM. THE TIP AND BALLOON SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. MANDREL RESISTANCE WAS ENCOUNTERED DUE TO THE PRESENCE OF SOLIDIFIED BLOOD WITHIN THE ENTIRE LENGTH OF THE INFLATION LUMEN, THEREFORE INDICATING THE DEVICE HAD BEEN USED IN VIVO. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(6) 2011. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE THE STENT DELIVERY SYSTEM WOULD NOT CROSS THE LESION. THE TARGET LESION BEING TREATED WAS LOCATED IN THE SEVERELY CALCIFIED UNKNOWN LOCATION. A 3.00X20MM PROMUS ELEMENT STENT WAS ADVANCED AND WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE. HOWEVER, THE RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY AND THE LESION LOCATION WAS IN THE LEFT ANTERIOR DESCENDING ARTERY.