FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 LONG

MDR report key: 2143202 · Received June 28, 2011

Report

Report Number
3005075853-2011-02669
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
June 7, 2011
Report Date
June 8, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE LONG60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRIC BYPASS PROCEDURE, ON THE INITIAL FIRING WITH A GREEN CARTRIDGE ON THE POUCH, THE STAPLE FORMATION WAS NOT COMPLETE. ON ONE ROW OF EACH SIDE THE STAPLES WERE MALFORMED. THE AREA WAS OVER SEWED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 LONG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1 ECR60G