FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE PRE P1.7
MDR report key: 2143179
·
Received June 28, 2011
Report
- Report Number
- 6000001-2011-09148
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- June 3, 2011
- Report Date
- June 3, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CONDITION OF ''SELECT'' BUTTON OF KEYPAD AT CHANNEL A WAS INOPERATIVE WAS CONFIRMED DUE TO FLUID INGRESS. THE KEYPAD WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
THIS IS A REPORT TO BAXTER (B)(4) OF A COLLEAGUE INFUSION PUMP IN WHICH THE ''SELECT'' BUTTON OF KEYPAD AT CHANNEL A WAS INOPERATIVE. THE REPORTED CONDITION OCCURRED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.48.92.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |