FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2143179 · Received June 28, 2011

Report

Report Number
6000001-2011-09148
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
June 3, 2011
Report Date
June 3, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CONDITION OF ''SELECT'' BUTTON OF KEYPAD AT CHANNEL A WAS INOPERATIVE WAS CONFIRMED DUE TO FLUID INGRESS. THE KEYPAD WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT TO BAXTER (B)(4) OF A COLLEAGUE INFUSION PUMP IN WHICH THE ''SELECT'' BUTTON OF KEYPAD AT CHANNEL A WAS INOPERATIVE. THE REPORTED CONDITION OCCURRED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.48.92.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1