ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2143144 · Received June 28, 2011

Report

Report Number
2531779-2011-04533
Event Type
Malfunction
Date Received
June 28, 2011
Report Date
June 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE UP ARROW BUTTON. MULTIPLE BUTTON PRESSES WERE REQUIRED BEFORE THE BUTTON WILL ENGAGE. THE BUTTONS ON THE KEYPAD DID NOT HAVE NORMAL SPRING BACK OR CLICK. THE KEYPAD WAS REMOVED TO CHECK CONDITION OF THE BUTTON CONTACTS. CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THAT THE PUMP WAS INTERMITTENTLY NOT RESPONDING TO UP BUTTON PRESSES. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 64 YR