FDA Adverse Event Injury Summary report: N

MINICAP

MDR report key: 2143118 · Received June 28, 2011

Report

Report Number
1423500-2011-08454
Event Type
Injury
Date Received
June 28, 2011
Date of Event
April 1, 2011
Report Date
June 6, 2011
Manufacturer
BAXTER HEALTHCARE - CLEVELAND
Product Code
KDI
PMA / PMN Number
K895631
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN AND PATIENTS DISCARD SUPPLIES AFTER EACH THERAPY, THE SAMPLE WAS NOT REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS REPORT 2 OF 2 INVOLVED IN THIS PERITONITIS EVENT.

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED FOR POTENTIALLY ASSOCIATED LOT NUMBER GD879924 WHICH REVEALED VOIDS WERE NOTED IN THE POUCH SEAL DURING MANUFACTURING. CORRECTIVE ACTIONS WERE IMPLEMENTED BY THE MANUFACTURING FACILITY AND A RE-INSPECTION OF THE LOT WAS PERFORMED WHICH REVEALED ALL AFFECTED UNITS WERE REMOVED PRIOR TO RELEASE OF THE LOT. THE CAUSE OF THE PERITONITIS WAS UNDETERMINED. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE (GPV) AND IS A SPONTANEOUS REPORT BY A CONSUMER FROM THE USA OF PERITONITIS IN A PATIENT COINCIDENT WITH DIANEAL PD4 AMBUFLEX AND DIANEAL PD4 ULTRABAG THERAPIES INTRAPERITONEALLY FOR PERITONEAL DIALYSIS (PD). DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED. ON AN UNREPORTED DATE, THE PATIENT EXPERIENCED PERITONITIS. THE CAUSE AND ORGANISM OF THE PERITONITIS WAS UNKNOWN. THE PATIENT WAS HOSPITALIZED FROM (B)(6) 2011 AND TREATMENT WAS NOT REPORTED. ON AN UNREPORTED DATE, THE PATIENT HAD RECOVERED FROM THE PERITONITIS. DIANEAL THERAPIES WERE ONGOING. AN OPINION OF CAUSALITY WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI BAXTER HEALTHCARE - CLEVELAND

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization HOMECHOICE| DIANEAL PD4 AMBUFLEX| DIANEAL PD4 ULTRABAG