FDA Adverse Event Malfunction Summary report: N

PROXIMATE ILS CURVED CIR STAPL

MDR report key: 2143001 · Received June 28, 2011

Report

Report Number
3005075853-2011-02664
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
June 1, 2011
Report Date
June 8, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DAMAGED ANCILLARY TROCAR. THE ANALYSIS RESULTS SHOWED THAT ONE ANVIL WAS RECEIVED WITH AN ANCILLARY TROCAR INSERTED. THE ANCILLARY TROCAR WAS REMOVED AND IT WAS NOTED THAT IT WAS MISALIGNED. IT WAS NOTED THAT THE ANCILLARY TROCAR WAS MOLDED INCORRECTLY RESULTING IN A DIMENSIONAL CHANGE CONSEQUENTLY INCREASING THE RESISTANCE FOR THE ATTACHMENT AND DETACHMENT OF THE ANCILLARY TROCAR. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. NO BATCH RECORD REVIEW COULD BE PERFORMED AS NO BATCH NOR LOT NUMBER WAS PROVIDED.

Description of Event or Problem · 1

ATTEMPTS FOR ADDITIONAL INFORMATION WERE COMPLETED ON (B)(4) 2011 AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A BARIATRIC PROCEDURE, THE BLUE PIN COULD ONLY STICK TOGETHER WITH EXTREME EFFORT. THE ADDITIONAL TROCAR IS "ABDOMINAL PROCESSED". NO FURTHER INFORMATION AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL FOLLOW UP IS BEING CONDUCTED DESPITE STATEMENT OF NO FURTHER INFORMATION IS AVAILABLE. IF ADDITIONAL DETAILS BECOME AVAILABLE A 3500 A SUPPLEMENTAL WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1