FDA Adverse Event Other Summary report: N

SPECTRA OPTIA COLLECT SET

MDR report key: 2142994 · Received June 21, 2011

Report

Report Number
1722028-2011-00184
Event Type
Other
Date Received
June 21, 2011
Date of Event
March 21, 2011
Report Date
May 30, 2011
Manufacturer
CARIDIANBCT
Product Code
LKN
PMA / PMN Number
K103090
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION: UPON VISUAL INSPECTION OF THE DISPOSABLE SET, THE FOLLOWING OBSERVATIONS WERE NOTED: THE UPPER BEARING WAS LACKING A WITNESS MARK TO INDICATE CORRECT LOADING (THE LOWER BEARING DOES HAVE SUCH A MARK). THE UPPER BEARING WAS WORN ADJACENT TO THE BRAID, WHICH CAN OCCUR WHEN THE BEARING IS INCORRECTLY LOADED IN TO THE HOLDER. THE BRAID WAS ALSO WORN ADJACENT TO THE BEARING, A LEAK WAS NOTED UNDER THE 4-LUMEN TUBING AT THIS LOCATION. AN IMAGE SUPPLIED WITH THIS COMPLAINT REPORT SHOWED THE BEARING TO BE IN THE CORRECT POSITION AFTER THE LEAK OCCURRED. UNFORTUNATELY, WE ARE UNABLE TO EXPLAIN THIS OBSERVATION. TRENDS SUGGEST THAT THE EVENT REPORTED IS RANDOM AND OCCURS AT A LOW LEVEL ON A GENERAL BASIS. ROOT CAUSE: BASED ON THE RETURNED DISPOSABLE, THE LOOP FAILURE IS BELIEVED TO HAVE BEEN CAUSED BY THE OPERATOR INCORRECTLY LOADING THE DISPOSABLE.

Description of Event or Problem · 1

DURING THE PROCEDURE, WITH ONE HOUR LEFT, THERE WAS AN ALARM FOR A CENTRIFUGE LEAK. THE COLLECTION WAS STOPPED. WHEN THE COVER WAS OPENED, IT WAS NOTED THAT THERE WAS SHEARING AND LEAKING OF A SMALL AMOUNT OF BLOOD IN THE MACHINE. ALSO, THE WEBBING ON THE TUBING WAS FRAYED, BUT NOT TWISTED. THE PATIENT INFORMATION IS UNAVAILABLE AT THIS TIME. PER INFORMATION AVAILABLE TO CARIDIANBCT, NO SAFETY ISSUE OCCURRED. CARIDIANBCT BECAME AWARE THAT THE CUSTOMER REPORTED THIS EVENT TO THEIR LOCAL AUTHORITIES ON (B)(4) 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRA OPTIA COLLECT SET SEPARATOR, AUTOMATED, BLOOD CELL AND PL LKN CARIDIANBCT 12S3114

Patients

Seq Age Sex Outcome Treatment
1 Other