LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-02660
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- June 3, 2011
- Report Date
- June 6, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE VISUAL INDICATOR WAS OVER-TRAVELED. WHEN THIS OCCURS, THE INDICATOR WHEEL MAY CONTINUE TO ROTATE AFTER INDICATION OF A COMPLETELY FIRED DEVICE. PLEASE NOTE THE OVER-TRAVEL OF THE INDICATOR WHEEL IS NOT RELATED TO THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY PROCEDURE, THE DEVICE WAS LOCKED OUT AND THE JAW BECAME NOT TO OPEN. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | H4385F |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |