FDA Adverse Event Injury Summary report: N

DIAMONDBACK 360 ORBITAL ATHERECTOMY SYSTEM

MDR report key: 2142695 · Received June 9, 2011

Report

Report Number
3004742232-2011-00017
Event Type
Injury
Date Received
June 9, 2011
Date of Event
May 13, 2011
Report Date
May 13, 2011
Manufacturer
CARDIOVASCULAR SYSTEMS INCORPORATED
Product Code
MCW
PMA / PMN Number
K071427
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE ANALYSIS: THE DEVICE WAS DISCARDED; THEREFORE, AN ANALYSIS OF THE ACTUAL COMPLAINT DEVICE IS NOT POSSIBLE AND THE ROOT CAUSE FOR THE REPORTED EVENT CANNOT BE CONFIRMED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ORBITAL ATHERECTOMY (OA) TREATMENT PROCEDURE, THE TIP OF THE VIPERWIRE WAS SHEARED OFF AND REMAINS IN THE PT'S BODY. THE LESION BEING TREATED WAS SEVERELY CALCIFIED, 70% STENOTIC. IT MEASURED APPROXIMATELY 13CM IN LENGTH AND WAS LOCATED IN THE ANTERIOR TIBIAL ARTERY. THE PHYSICIAN USED A 6F INTRODUCER SHEATH TO CROSS THE TARGET LESION. USING THE DIAMONDBACK 1.5 CLASSIC CROWN OAD, THE PHYSICIAN PERFORMED TWO RUNS AT LOW SPEED (80K RPMS) LASTING 25SECS EACH. AFTER 15SECS DURING THE THIRD RUN AT 80K RPMS, THE DIAMONDBACK OAD CROWN WAS PUSHED OVER THE TIP OF THE VIPERWIRE CAUSING THE TIP OF THE WIRE TO BE SHAVED OFF IN THE PT'S AT VESSEL. IT WAS NOTED THAT HALF-WAY THROUGH THIS RUN THE FLUORO TURNED OFF PREVENTING VISIBILITY IN THE TREATMENT REGION. THE CAUSE OF THE LOSS OF FLUORO IS UNK. THE DEVICE MADE A HIGH-PITCHED NOISE AND THE INTERVENTION WAS IMMEDIATELY HALTED. THE VASCULAR SURGEON WAS CONSULTED AND IT WAS DETERMINED THAT IT WAS NOT NECESSARY TO REMOVE THE TIP OF THE WIRE AS THE PT'S LESIONS WERE IMPROVED AND SHE HAD A PALPABLE PULSE IN THE AT WHICH WAS NON-EXISTENT PRIOR TO TREATMENT. ADDITIONAL INFO HAS BEEN REQUESTED REGARDING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIAMONDBACK 360 ORBITAL ATHERECTOMY SYSTEM MCW CARDIOVASCULAR SYSTEMS INCORPORATED VPR-GW-14

Patients

Seq Age Sex Outcome Treatment
1 Other