Malfunction
Summary report: N
CENTURY BED
MDR report key: 2142622
·
Received June 21, 2011
Report
- Report Number
- 1824206-2011-03364
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- June 2, 2011
- Report Date
- June 2, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN TIGHTENED ALL OF THE GROUND STRAPS TO RESOLVE THE LOSS OF GROUND ISSUE.
Description of Event or Problem · 1
THE TECHNICIAN FOUND THE GROUND RESISTANCE READING WAS HIGH WHILE PERFORMING AN ELECTRICAL SAFETY TEST DURING A PREVENTIVE MAINTENANCE SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTURY BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1400 |