FDA Adverse Event Malfunction Summary report: N

CENTURY BED

MDR report key: 2142622 · Received June 21, 2011

Report

Report Number
1824206-2011-03364
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
June 2, 2011
Report Date
June 2, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN TIGHTENED ALL OF THE GROUND STRAPS TO RESOLVE THE LOSS OF GROUND ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN FOUND THE GROUND RESISTANCE READING WAS HIGH WHILE PERFORMING AN ELECTRICAL SAFETY TEST DURING A PREVENTIVE MAINTENANCE SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1400

Patients

Seq Age Sex Outcome Treatment
1