FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOGUARD

MDR report key: 21426189 · Received February 20, 2025

Report

Report Number
1710034-2025-00290
Event Type
Malfunction
Date Received
February 20, 2025
Date of Event
February 6, 2025
Report Date
June 16, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903814237
PMA / PMN Number
K201075
Removal / Correction Number
MDS-25-5274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

EVENT DATE UPDATED IN B TAB INVESTIGATION RESULTS: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED WERE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. HOWEVER, THE REPORTED DEFECT OF NEEDLE RETRACTION SLOW IS CONFIRMED SINCE A TREND HAS BEEN IDENTIFIED FOR THE REPORTED FAILURE MODE AND CORRECTIVE ACTIONS HAVE BEEN INITIATED TO INVESTIGATE THIS TYPE OF INCIDENT AND IDENTIFY THE ROOT CAUSE.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION OF FA#.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Description of Event or Problem · 0

EVENT DATE IS UNKNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOGUARD NEEDLE WAS SLOW TO RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MULTIPLE OF BD INSYTE AUTOGUARD IV'S ARE MALFUNCTIONING. IV NEEDLE IS NOT RETRACTING AS IT SUPPOSED TO. WHEN YOU PUSH THE BUTTON TO RETRACT THE NEEDLE IT DOES NOT RETRACT. ONCE YOU PULL THE NEEDLE OUT OF THE CATHETER OUT OF THE PT AND THEN HIT THE BUTTON IT WILL RETRACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
394700 BD INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4229661 00382903814237

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown