FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2142603 · Received June 21, 2011

Report

Report Number
2032227-2011-01522
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
June 6, 2011
Report Date
June 6, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH A BLANK DISPLAY DUE TO CORROSION ON THE LIQUID CRYSTAL DISPLAY BOARD. UNABLE TO VERIFY THE FROZEN DISPLAY DUE TO THE BLANK DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY AND UNRESPONSIVE BUTTONS. TROUBLESHOOTING WAS PERFORMED, AND THE ISSUE WAS NOT RESOLVED. THE TIME ON THE SCREEN WAS NOT ADVANCING. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722RNAS

Patients

Seq Age Sex Outcome Treatment
1 68 YR