FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 2142584 · Received June 21, 2011

Report

Report Number
1824206-2011-03373
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
June 2, 2011
Report Date
June 2, 2011
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE DETENT WAS CRACKED AND BROKEN IN HALF. HE REPLACED THE BRAKE DETENT TO REPAIR THE BED. (B)(4) IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT MECHANISM. THIS IS A POST MOD FAILURE.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BRAKE PEDAL WOULD GO INTO THE BRAKE POSITION BUT THE BRAKE CASTER WOULD NOT HOLD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM INC. 3700B

Patients

Seq Age Sex Outcome Treatment
1