FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2142567 · Received June 21, 2011

Report

Report Number
1824206-2011-03351
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
May 31, 2011
Report Date
May 31, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT'S MAINTENANCE STATED HE CHECKED THE CPR HANDLE AND IT SEEMED LIKE THE BASE MIDDLE COVER MAY HAVE BEEN PUSHED UP AGAINST THE CPR LEVER AND CAUSING IT TO STICK OPEN. HE RE-ADJUSTED THE BASE MIDDLE COVER TO REPAIR BE BED.

Description of Event or Problem · 1

THE ACCOUNT'S MAINTENANCE STATED THE HEAD SECTION WILL LOWER BACK DOWN AFTER HE LETS GO OF THE HEAD UP BUTTON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1