FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2142567
·
Received June 21, 2011
Report
- Report Number
- 1824206-2011-03351
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- May 31, 2011
- Report Date
- May 31, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT'S MAINTENANCE STATED HE CHECKED THE CPR HANDLE AND IT SEEMED LIKE THE BASE MIDDLE COVER MAY HAVE BEEN PUSHED UP AGAINST THE CPR LEVER AND CAUSING IT TO STICK OPEN. HE RE-ADJUSTED THE BASE MIDDLE COVER TO REPAIR BE BED.
Description of Event or Problem · 1
THE ACCOUNT'S MAINTENANCE STATED THE HEAD SECTION WILL LOWER BACK DOWN AFTER HE LETS GO OF THE HEAD UP BUTTON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |