FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2142503 · Received June 28, 2011

Report

Report Number
1423500-2011-08441
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
May 18, 2011
Report Date
June 6, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE PASSED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION), FUNCTIONAL AND ELECTRICAL TESTS AND WAS FUNCTIONING WITHIN SPECIFICATION. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RTB-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA PERITONEAL VOLUME(IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE IIPV OCCURRED ON (B)(6) 2011 DURING CYCLE 2 . THE PROGRAMMED FILL VOLUME WAS 2000 ML AND DRAIN VOLUME WAS 3344 ML. THIS MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THIS IS REPORT 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 54 YR