HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-08434
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- May 15, 2011
- Report Date
- June 6, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE PASSED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION), FUNCTIONAL AND ELECTRICAL TESTS AND WAS FUNCTIONING WITHIN SPECIFICATION.THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RTB-CAPA-(B)(4).
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA PERITONEAL VOLUME(IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE IIPV OCCURRED ON (B)(6) 2011 DURING CYCLE 2 . THE PROGRAMMED FILL VOLUME WAS 2000 ML AND DRAIN VOLUME WAS 3262 ML . THIS MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THIS IS REPORT 1 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |