ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-04529
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 14, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON 05/31/2011. RECALL 2531779-03/24/2010-003-R. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION AND THE FORCE SENSOR FLEX CONNECTOR WAS FOUND TO BE PARTIALLY TORN. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE PUMP WAS OPERATING WITHIN SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |