FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2142374 · Received June 28, 2011

Report

Report Number
2531779-2011-04528
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
April 13, 2011
Report Date
May 6, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON 06/01/2011. (B)(6). RECALL 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED. UNRELATED TO THE COMPLAINT, THE PUMP'S BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THE BATTERY CAP WAS FOUND TO BE STRIPPED, AND WOULD NOT MAINTAIN POWER. FURTHER TESTING WAS COMPLETED WITH A TEST BATTERY CAP.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1