FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2142351 · Received June 28, 2011

Report

Report Number
2531779-2011-04526
Event Type
Malfunction
Date Received
June 28, 2011
Report Date
June 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. INSPECTION REVEALED THAT THE BUTTON SYMBOLS ARE WORN. EVALUATION CONFIRMED THAT THE OK KEYPAD BUTTON IS INTERMITTENTLY RESPONSIVE. SCROLLING WAS NOT OBSERVED DURING TESTING. DURING INVESTIGATION, IT WAS NOTED THAT NONE OF THE KEYPAD BUTTONS HAVE NORMAL SPRING BACK, AND THE OK BUTTON CONTACT WAS FOUND TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE OK KEYPAD BUTTON BECAME INTERMITTENTLY UNRESPONSIVE LAST WEEK. SHE NOTED THAT THE OK BUTTON HAD A DELAYED RESPONSE, AND STATED THAT WHEN IT DOES RESPOND, THE PUMP SCROLLS THROUGH VARIOUS SCREENS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 48 YR