FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2142313
·
Received June 28, 2011
Report
- Report Number
- 2531779-2011-04525
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Report Date
- April 30, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2011. CORRECTION NUMBER: 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED AND THERE WAS CONTAMINATION FOUND ON THE FORCE SENSOR.
Additional Manufacturer Narrative · 1
(B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |