FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2142284 · Received June 28, 2011

Report

Report Number
2531779-2011-04523
Event Type
Malfunction
Date Received
June 28, 2011
Report Date
June 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN-ARROW AND OK BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THE UP-ARROW AND CONTRAST BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE. THE UP-ARROW AND DOWN-ARROW KEY CONTACTS WERE FOUND TO BE MISALIGNED. IT WAS FOUND THAT ALL KEY CONTACTS DO NOT SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. THE DHR REVIEW CONFIRMED THE PUMP WAS WITHIN SPECIFICATION WITH NO DISCREPANCIES IDENTIFIED, SOFTWARE VERSION D1E, MFG 2009-04.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE KEYPAD HAS A DELAYED RESPONSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 66 YR