COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-09125
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- January 1, 2011
- Report Date
- January 5, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF "BAD DISPLAY" WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS NOT IDENTIFIED. NO FURTHER TESTING HAS BEEN COMPLETED AT THIS TIME AND NO REPAIRS HAVE BEEN PERFORMED AS THE REFERENCED DEVICE HAS BEEN REMOVED FROM SERVICE. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A "BAD DISPLAY." IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |