STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-02941
- Event Type
- Injury
- Date Received
- February 20, 2025
- Date of Event
- January 14, 2025
- Report Date
- October 19, 2025
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: A2;
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON OCTOBER 29, 2025, WITH LOT NUMBER 1851162. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION (IF APPLICABLE). PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ DEFLATION: OBSERVED MULTIPLE OPENINGS THROUGH MICROSCOPIC INSPECTION 1 OPENING ASSESSED AS SURGICAL DAMAGE AND 1 OPENING ASSESSED AS FOLD FLAW OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES, NON-PENETRATING NICKS AND WEAR ABRASION) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B.5., D6B., H.6.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED.
DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1038148 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1851162 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Female | Required Intervention |