FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2142267 · Received June 28, 2011

Report

Report Number
2531779-2011-04521
Event Type
Malfunction
Date Received
June 28, 2011
Report Date
June 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. EVALUATION REVEALED THAT THE KEYPAD WAS TORN. A TORN KEYPAD WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD SHOULD BE CLEARLY VISIBLE AND WARNS THE PATIENT TO DISCONTINUE USING THE PUMP. ADDITIONALLY MOISTURE WAS OBSERVED UNDER THE DISPLAY. THE PUMP WAS UNABLE TO POWER ON. THE PUMP WAS OPENED AND MOISTURE WAS OBSERVED IN THE CIRCUIT BOARD.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE RUBBER ON THE OK BUTTON STARTED PEELING ONE WEEK AGO. SHE ALSO REPORTED THAT SHE THINKS THE PUMP GOT WET AT SCHOOL TODAY AND NOW HAS NO POWER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 7 YR