FDA Adverse Event Malfunction Summary report: N

ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM

MDR report key: 2142263 · Received June 28, 2011

Report

Report Number
2134265-2011-02612
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
June 10, 2011
Report Date
June 14, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
NIQ
PMA / PMN Number
P100023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: RETURNED PRODUCT CONSISTED OF AN ION STENT DELIVERY SYSTEM (SDS) WITH THE SHELF BOX. THE BALLOON WAS TIGHTLY FOLDED WITH THE STENT IMPLANT SECURE BETWEEN THE MARKERBANDS. FOUR STRUTS IN THE FIRST DISTAL ROW WERE STRETCHED DISTALLY. THERE WAS NO OTHER DAMAGE TO THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). WHEN THE PHYSICIAN PULLED THE 3.00X12MM ION STENT DELIVERY SYSTEM (SDS) OUT OF THE PACKAGE THE STENT LOOKED DAMAGED. THE DEVICE WAS PUT TO THE SIDE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). WHEN THE PHYSICIAN PULLED THE 3.00X12MM ION STENT DELIVERY SYSTEM (SDS) OUT OF THE PACKAGE THE STENT LOOKED DAMAGED. THE DEVICE WAS PUT TO THE SIDE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - MAPLE GROVE H7493902312300 14265409

Patients

Seq Age Sex Outcome Treatment
1