FDA Adverse Event Malfunction Summary report: N

MCA MED APPLIER

MDR report key: 2142241 · Received June 28, 2011

Report

Report Number
3005075853-2011-02614
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
June 7, 2011
Report Date
June 8, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDO
PMA / PMN Number
K820837
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: NO CLIPS FELL INTO THE PATIENT'S BODY.

Additional Manufacturer Narrative · 1

(B)(4). HOOP SPRING THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE TRIGGER COULD NOT BE COMPLETELY ACTIVATED DUE TO THE FIRING MECHANISM WAS JAMMED. IN ORDER TO EVALUATE THE CONDITION OF THE DEVICE'S INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE HOOP SPRING WAS FOUND OUT OF ITS INTENDED POSITION, JAMMING THE FIRING MECHANISM AND MAKING THE ANTIBACKUP FEATURE NON-FUNCTIONAL. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE FOUND CONDITION OF THE HOOP SPRING. IT IS KNOWN FROM THE HISTORY OF THE DEVICE THAT THE ANTIBACKUP FAILURE MAY LEAD TO DROPPING CLIPS. IN ORDER TO EVALUATE THE PROPER ASSEMBLY OF THE HOOP SPRING, IT WAS REASSEMBLED. UPON FIRING, THE DEVICE PERFORMED AS INTENDED; THE CLIPS WERE FED AND PROPERLY FORMED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN CORONARY ARTERY BYPASS GRAFT, THE CLIP FELL OUT FROM THE JAW AT THE FIRST FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MCA MED APPLIER CLIP, IMPLANTABLE GDO ETHICON ENDO-SURGERY, LLC. NA UNK