COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-09113
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- May 31, 2011
- Report Date
- June 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH DIAGNOSTIC FAILURE CODE 810:11 WAS CONFIRMED BUT NOT DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE WAS DETERMINED TO BE AN OUT OF CALIBRATION AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB). THE AIL PCB WAS RECALIBRATED AND VERIFIED TO CORRECT THIS CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO INFORMATION REGARDING PATIENT INVOLVEMENT, PATIENT INJURY, OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |