FDA Adverse Event Summary report: N

COBAS TAQSCREEN MPX TEST

MDR report key: 2142149 · Received June 28, 2011

Report

Report Number
2243471-2011-00063
Date Received
June 28, 2011
Date of Event
March 1, 2011
Report Date
August 30, 2011
Manufacturer
ROCHE MOLECULAR SYSTEMS
Product Code
MKT
PMA / PMN Number
BL125255
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A DEFINITIVE CONCLUSION CANNOT BE DRAWN AT THIS TIME, AS THE INVESTIGATION INTO THIS ISSUE IS ONGOING. THE OUTCOME OF THE INVESTIGATION WILL BE COMMUNICATED THROUGH A FOLLOW-UP REPORT. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4): RESULT - DEVICE PERFORMED ACCORDING TO SPECIFICATIONS. CONCLUSION - NO FAILURE DETECTED AND PRODUCT IS WITHIN SPECIFICATION. THE CUSTOMER REPORTED THAT THEY OBTAINED A NON-REACTIVE RESULT FROM A DONOR WITH THE COBAS TAQSCREEN MPX TEST BUT THE DONOR SPECIMEN WAS SEROLOGY (B)(4). TWO COLLECTIONS FROM THE SAME DONOR WERE TESTED AND BOTH GENERATED NON-REACTIVE MPX RESULTS. THE SECOND COLLECTION WAS ALSO TESTED BY THE CUSTOMER WITH THE COBAS AMPLIPREP / COBAS (CAP/CTM) TAQMAN HBV TEST AND GENERATED A RESULT OF TARGET NOT DETECTED (TND), INDICATING THAT THE SAMPLE LIKELY HAS A TITER BELOW THE LIMIT OF DETECTION (LOD) OF 12 IU/ML OF THE CAP/CTM HBV TEST. NO SAMPLES WERE AVAILABLE FROM THE CUSTOMER FOR FURTHER EVALUATION THE DONATION WAS BLOCKED FROM THE BLOOD SUPPLY. BASED ON THE INFORMATION AVAILABLE, NO PRODUCT NON-CONFORMANCE WAS IDENTIFIED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER SITE IN (B)(4) REPORTED THAT A PATIENT SAMPLE GENERATED A FALSE (B)(6) TEST RESULT, FOR HBV, WHEN TESTED WITH THE COBAS TAQSCREEN MPX TEST. SPECIFICALLY, THE CUSTOMER REPORTED THAT THE DONOR SPECIMEN WAS SEROLOGY (B)(6) AND GENERATED (B)(6) TEST RESULTS WITH THE COBAS TAQSCREEN MPX TEST. THE PATIENT SAMPLE WAS TESTED FOR SEROLOGY MULTIPLE TIMES WITH THE ARCHITECT (ABBOTT) AND MUREX (DIASORIN) TESTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS TAQSCREEN MPX TEST HEPATITIS VIRAL B DNA DETECTION MKT ROCHE MOLECULAR SYSTEMS