PENUMBRA COIL 400 DETACHMENT HANDLE
Report
- Report Number
- 3005168196-2011-00228
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- April 28, 2011
- Report Date
- April 28, 2011
- Manufacturer
- PENUMBRA, INC.
- Product Code
- HCG
- PMA / PMN Number
- K103305
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
RESULTS: THERE WERE NO NOTICEABLE DEFECTS WITH EITHER HANDLE. THE UNITS WERE TESTED FOR PULL FORCE AND THROW DISTANCE ON THE PRODUCTION TEST FIXTURE. BOTH UNITS PASS THE APPLIED TESTS. THIS ALSO CONFIRMS THE HANDLE FUNNEL IS WITHIN SPECIFICATION. BOTH HANDLES ARE FULLY FUNCTIONAL. CONCLUSION: THE HANDLES WERE RETURNED AS PART OF A GENERAL COMPLAINT ABOUT THE DIFFICULTY OF DETACHING COILS IN THIS PATIENT. THE HANDLES DID NOT CONTRIBUTE TO THE DETACHMENT ISSUES. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN OR QUALITY CONCERNS.
THE PHYSICIAN WAS ATTEMPTING TO COIL AN UNRUPTURED, RIGHT, 11 MM BASILAR ARTERY ANEURYSM IN A PATIENT WITH VERY TORTUOUS ANATOMY. THE PHYSICIAN MADE THREE ATTEMPTS TO DETACH THE FIRST COIL USING A PENUMBRA COIL 400 DETACHMENT HANDLE, EACH TIME UNSUCCESSFULLY. THE PHYSICIAN WITHDREW THE COIL, AND ATTEMPTED TO PLACE ANOTHER COIL WITH A NEW DETACHMENT HANDLE. AFTER SEVERAL UNSUCCESSFUL TRIES WITH THE HANDLE, THE PHYSICIAN TRIED TO MANUALLY SEPARATE THE PULL WIRE BY HAND. THIS ATTEMPT ALSO FAILED TO DEPLOY THE COIL. THE PHYSICIAN COMMENTED THAT AS HE PULLED ON THE PULL WIRE, HE FELT A STRETCHINESS IN THE WIRE AND A REFLUX BACK ON EACH PULL LIKE A RUBBER BAND. THE PHYSICIAN THEN DECIDED TO REMOVE THE COIL, WHICH RESULTED IN THE COIL DETACHING INSIDE THE PENUMBRA PX400 MICROCATHETER. HE THEN DECIDED TO PUSH THE COIL BACK INTO THE ANEURYSM WITH THE PUSHER. THE COIL WAS SUCCESSFULLY ABLE TO BE REDEPLOYED BACK INTO THE ANEURYSM. THE PHYSICIAN REMOVED THE PX400 AND USED ANOTHER PRODUCT TO COMPLETE THE COILING. THE PATIENT WAS ALERT AND INTACT AFTER THE PROCEDURE. THIS MDR IS ASSOCIATED WITH MDRS 3005168196-2011-00229 THROUGH 3005168196-2011-00232.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PENUMBRA COIL 400 DETACHMENT HANDLE | HCG, KRD | HCG | PENUMBRA, INC. | F18238 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |