ON-Q C-BLOC SAF/PCA
Report
- Report Number
- 2026095-2011-00149
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- May 6, 2011
- Report Date
- May 6, 2011
- Manufacturer
- I-FLOW CORP.
- Product Code
- MEB
- PMA / PMN Number
- K063530
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NURSE
Narratives
METHOD: SAMPLE HAS NOT YET BEEN REC'D, BUT WAS REPORTED TO BE AVAILABLE. A REVIEW OF THE DEVICE HISTORY RECORDS (DHR) WAS CONDUCTED FOR THE LOT NUMBER AND INCIDENT REPORTED. THE DEVICE PASSED ALL MFG SPECIFICATIONS PRIOR TO RELEASE. THE DIRECTIONS FOR USE (DFU) (PN 1306085, REV. B) STATES: WARNING: IF THE BUTTON DOES NOT POP BACK UP WITHIN 30 MINUTES, CHECK POSITION OF ORANGE REFILL INDICATOR. IF INDICATOR IS IN THE LOWERMOST POSITION, CLOSE CLAMP. IF BUTTON POPS BACK UP WITHIN 10 MINUTES, OPEN CLAMP AND CONTINUE WITH INFUSION. IF BUTTON REMAINS STUCK, IT MAY RESULT IN A SIGNIFICANT INCREASE IN FLOW RATE TO PT. KEEP CLAMP CLOSED. PUMP SHOULD BE REPLACED IF APPROPRIATE. IF INDICATOR REMAINS IN THE UPPERMOST POSITION, SOMETHING MAY BE IMPEDING THE FLOW. CHECK FOR TUBING KINKS, CLOSED CLAMP OR PATENCY OF CONNECTED DEVICES SUCH AS CATHETER OR UNVENTED FILTER (VERIFY PATENCY) ACCORDING TO YOUR STANDARD PROTOCOL." RESULTS: WITHOUT THE ACTUAL PRODUCT, AN ANALYSIS CANNOT BE CONDUCTED. CONCLUSION: THE SAMPLE WILL BE EVALUATED WHEN REC'D AND A F/U UP REPORT WILL BE FILED.
DRUG/DILUENT: ROPIVACAINE 0.2%. FILL VOLUME: 400 AND FLOW RATE: 6ML/HR. PROCEDURE: OPEN REDUCTION INTERNAL FIXATION, LEFT ANKLE FRACTURE. CATHPLACE: POPLITEAL BLOCK. BOLUS BUTTON KEPT STICKING AND WOULD NOT POP BACK UP. WHEN REP WENT TO CHECK ON THE PUMP, THE BOLUS SEEMED TO BE WORKING FINE. PUMP WAS REPLACED FOR PT. NO ADVERSE EVENT REPORTED. DATE OF EVENT: (B)(6) 2011.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ON-Q C-BLOC SAF/PCA | ELASTOMERIC PUMP | MEB | I-FLOW CORP. | CB005 | 112157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |