FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 2142066 · Received May 25, 2011

Report

Report Number
2936999-2011-00350
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
April 28, 2011
Report Date
April 28, 2011
Manufacturer
COVIDIEN/FORMERLY TYCO HEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE EVAL FOUND THE TEMPERATURE CONTROL OUT OF ORDER, THE ELECTRONIC PCB AND THERMISTOR WERE BROKEN. THE KEYPAD WAS WORN OUT. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS K020604.

Description of Event or Problem · 1

IT WAS REPORTED BY COVIDIEN TECHNICIAN THAT THE WARMTOUCH TEMPERATURE WAS TOO HIGH. THE UNIT WOULD SHORT HEAT TO 53 DEGREESC IN ALL HEATING LEVELS. THE UNIT DID NOT SHUT DOWN OR GIVE AN ALARM. THERE WAS NO PT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARMTOUCH COMPACT WARMING UNIT CONVECTIVE AIR WARMER DWJ COVIDIEN/FORMERLY TYCO HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1