FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2141923 · Received June 28, 2011

Report

Report Number
2050012-2011-02652
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
May 28, 2011
Report Date
May 28, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER FOUND BUBBLE IN THE ION-SELECTIVE ELECTRODE (ISE) BUFFER AND OBSERVED THAT THE FITTING ON THE REAGENT BOTTLE WAS LOOSE. A BCI FIELD SERVICE ENGINEER (FSE) FOUND LINE #24 PINCHED BETWEEN THE FLOW CELL BRACKET AND THE CHASSIS. FSE ELECTROLYTE INJECTION CUP (EIC) VALVE MOD 10874 AND COMPLETED THE ISE CALIBRATION HEALTH CHECK AND PRECISION HEALTH CHECK. FSE RECEIVED A WASTE SUMP FULL ERROR WHEN PERFORMED THE QC TEST, WHICH SHUT THE ANALYZER DOWN. FSE CLEANED AND EMPTIED THE SUMP AND THE VALVE. FSE PRIMED THE INSTRUMENT 20 TIMES. FSE TESTED QC AND OBSERVED CARSOUEL MOTION ERROR. FSE RESTARTED THE INSTRUMENT AND COMPLETED THE QC TEST. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT THAT THE ELECTROLYTES FAILED CALIBRATION. SODIUM SAMPLE REFERENCES WERE INCONSISTENT. NO INJURY OR EXPOSURES WERE REPORTED.